beautypg.com

2 weighing, Switching on the scale – Seca 955 User Manual

Page 17

background image

Operation • 17

English

6.2

Weighing

CAUTION!

Patients can be injured if they fall

People with limited mobility can fall if they try to
sit down on the scale.
– Make sure that the scale’s wheel brakes are

set.

– Assist people with limited mobility when they

are sitting down.

– Leave the footrests folded up and pushed in

until the patient is sitting on the scale.

CAUTION!

Patients can be injured if their limbs get pinched

The patient’s hands and arms can be pinched
between the seat and the lower frame or
between the seat and the armrests.
– Make sure that the patient’s hands and arms

are on the armrests or in his lap during the
weighing process.

ATTENTION!

Incorrect measurement due to force shunts

If accessories (e.g. the power pack pouch) or
articles of clothing get caught between the seat
and the frame, the weight will not be measured
correctly.
If the patient’s feet are not on the footrests, the
weight will not be measured correctly.
– Make sure that there are no accessories or

articles of clothing between the seat and
frame.

– Make sure that the armrests are folded down

during the weighing process.

– Make sure that the patient’s feet are on the

footrests during the weighing process.

Switching on the scale



Press the Start key.
The display will read

se[A

and then quickly run

through all elements of the display.
The scales are ready for operation when the display
reads

0.00

.

If the device is being operated with a power supply
unit, the symbol

will appear in the display.