beautypg.com

Tare off additional weight (tare), Continuous display of measured result (hold) – Seca 955 User Manual

Page 19

background image

Operation • 19

English

Tare off additional

weight (TARE)

You can use the TARE function to prevent the weighing
result from being affected by an additional weight (e.g.
a towel or pad on the weighing surface).

ATTENTION!

Incorrect measurement with force shunt

If an additional weight (e.g. large towel) is in
contact with the surface on which the scales
are standing, the weight will not be measured
correctly.
– Make sure that any additional weight is only

placed on the weighing surface of the scales.

1. Switch the scale on.

2. Place the additional weight on the scales.

3. Hold down the arrow key (hold/tare) until "NET"

appears in the display.

4. Wait until the display stops flashing and is replaced

by

0.00

.

5. Weigh the patient as described in section “Start

weighing”.

6. Read off the measured result.

The additional weight has been automatically
deducted.

7. To disable the TARE function, press the arrow key

(hold/tare) until "NET" no longer appears in the
display or switch off the scales.

NOTE:
The maximum weight that can be displayed is
reduced by the weight of the objects already
placed on the scales.

Continuous display of

measured

result (HOLD)

If you activate the HOLD function, the weighing value
will continue to be displayed after the load has been
removed. This allows you to attend to the patient
before noting down the weight.

1. Check that there is no load on the scales.

2. Switch the scale on.

3. Weigh the patient as described in section “Start

weighing”.

4. Briefly press the arrow key (hold/tare).

tare

tare

tare