beautypg.com

Seca 955 User Manual

Page 8

background image

8

– The rear casters can be locked if you

push down the two locking levers on

the casters. This device should be used

when the patient is sitting down on the

scale or getting up from it.

– Once the person who is to be weighed

is seated, the foot-rests are tilted down-
wards. The patient can now place his or

her feet on the foot-rests.

Checking that the contents of the calibration counter are

correct

This seca scale is calibrated. Calibration may be performed only by authorised centres.

To ensure this, the scale is fitted with a calibration counter which records any change to

data relevant to calibration. If you want to check whether the scale is properly calibrated,

proceed as follows:
– If necessary, switch off the scale.
– Keep any key depressed and start the

scale.
The current contents of the calibration

counter are shown on the display for a

few seconds.

– Compare the output contents of the cal-

ibration counter with the number shown

on the calibration counter sticker.
For valid calibration, both numbers must

match. If sticker and calibration counter

do not match, then re-calibration is re-
quired. Contact your service agent or

seca Customer Service.

Locking lever

Both numbers must match.