8. Servicing/recalibration

Have the scale recalibrated by authorised persons in accordance with national legal re-
quirements. The year of initial calibration can be found behind the CE symbol on the rating

plate above the number of the appointed office, 0109 (Hessen Weights & Measures Of-
fice). Recalibration is essential if one or more security marks are damaged or the calibra-
tion counter does not match the number on the applicable calibration counter mark (cf.
page 8).
We recommend having your scale serviced by your local service agent before having it

recalibrated. seca Customer Service will be pleased to give you further assistance.

9. Technical data

Maximum load

200 kg

Minimum load

2 kg

Graduations

100 g

Dimensions
height:

920 mm

width:

565 mm

depth:

810 mm

Weight

approx. 24.5 kg

Temperature range

+10 °C to +40 °C

Power supply

rechargeable batteries/mains unit

Power consumption

approx. 25 mA

Calibration as per Directive

2009/23 EC

approved for medical use, Class III

Medical product as per Directive

93/42/EEC

Class I

EN 60 601-1:
Insulated equipment, Protection Class II:
medical electrical equipment type B:
Accuracy at first calibration
2 kg to 50 kg:

± 50 g

50 kg to 200 kg:

± 100 g

Functions

Tare, Hold, BMI