8. Servicing/recalibration
Have the scale recalibrated by authorised persons in accordance with national legal re-
quirements. The year of initial calibration can be found behind the CE symbol on the rating
plate above the number of the appointed office, 0109 (Hessen Weights & Measures Of-
fice). Recalibration is essential if one or more security marks are damaged or the calibra-
tion counter does not match the number on the applicable calibration counter mark (cf.
page 8).
We recommend having your scale serviced by your local service agent before having it
recalibrated. seca Customer Service will be pleased to give you further assistance.
9. Technical data
Maximum load
200 kg
Minimum load
2 kg
Graduations
100 g
Dimensions
height:
920 mm
width:
565 mm
depth:
810 mm
Weight
approx. 24.5 kg
Temperature range
+10 °C to +40 °C
Power supply
rechargeable batteries/mains unit
Power consumption
approx. 25 mA
Calibration as per Directive
2009/23 EC
approved for medical use, Class III
Medical product as per Directive
93/42/EEC
Class I
EN 60 601-1:
Insulated equipment, Protection Class II:
medical electrical equipment type B:
Accuracy at first calibration
2 kg to 50 kg:
± 50 g
50 kg to 200 kg:
± 100 g
Functions
Tare, Hold, BMI