12

6. Cleaning

Clean the seat and the housing as required using a domestic cleaning agent or commer-
cially-available disinfectant. Follow the manufacturer’s instructions.
Under no circumstances use abrasive or acid cleaners, white spirit, benzene or the like for

cleaning. Such substances can damage the high-quality surfaces.

7. What do I do if…

… no weight display comes on?
– Is the scale switched on?
– Check the mains connection and the connection of the battery pack.
– Switch the scale off for a few seconds. This can be done by opening the plug connect-

ed to the rechargeable battery block and unplugging the mains supply from the scale.

0.0

does not appear before weighing?

– Switch off the scale with the Start key and start the scale again – there must not be

any load on the scale – and only its feet should be in contact with the floor.

… one segment is illuminated constantly or not at all?
– The relevant segment has a fault. Inform Servicing.

… the display shows

?

– Battery voltage is dropping. Charge the rechargeable batteries in the next few days.

bAtt

appears in the display?

– The rechargeable battery is so severely discharged that weighing is no longer possible.

Recharge the battery.

St0p

appears in the display?

– Maximum load has been exceeded.

… the display flashes?
– If you have just activated the Hold or Tare function, please wait until the scale has de-

tected a steady weight value. The display will then stop flashing.

– If there has been no activation of a function beforehand, remove the load from the scale

and wait until

0.0

is displayed, then weigh again.

… the display

appears?

– The ambient temperature of the scale is too high or too low. Place the scale in an

ambient temperature between +10 °C and +40 °C. Wait about 15 minutes for the

scale to adapt to the ambient temperature and then weigh again.

… the display

E

and a number appear?

– Switch off the scale with the Start key and start the scale again. The scale will then

work normally again.

If this is not the case, disconnect the power supply by briefly removing the batteries. If

this measure is equally unsuccessful, inform the Service department.