LumaSense Technologies M305 User Manual
Page 16
M305 Manual
Installation and Operation 16
Typical Radiance temperature calibration look-up table (LUT) used for this M305
To gain access to the custom look-up table:
Refer to the 4 elliptical buttons on the face of the controller and use this procedure.
1. Press and Hold 1 until ACCESS appears. If Access Level and Level 1 appears, skip to
step 3.
2. Press 4 until Level 3 appears.
3. Press 4 until 3 appears. Wait and "pass" appears.
4. Press 1 until Lin16 appears.
5. Press 2. until display shows the Output # that corresponds to the calibration point
shown in the listing on the previous page.
6. Use UP/DWN buttons (3 and 4) to dial in the measured radiance temperature from a
freezing point transfer device or other radiance standard. Adjust Out# only!
1
2
3
4
See also other documents in the category LumaSense Technologies Tools:
- IN 510-N (38 pages)
- M67S (52 pages)
- IGA 740-LO (28 pages)
- IGA 740 (27 pages)
- IN 3000 (4 pages)
- ISR 320 (46 pages)
- IGA 320/23-LO (44 pages)
- IGA 320 (34 pages)
- IGA 310 (21 pages)
- IS 50/067-LO plus (36 pages)
- IS 6-TV Advanced (25 pages)
- IN 6/78-L (38 pages)
- ISR 6 Advanced (50 pages)
- IGA 6/23 Advanced (46 pages)
- IGA 6 Advanced (44 pages)
- IGA 8 pro (32 pages)
- IS 12-Al (38 pages)
- IGA 12 (40 pages)
- M8100PM-EXP (78 pages)
- MCS640 Manual (26 pages)
- MCS640 Quick Start Guide (4 pages)
- MC320 Manual (32 pages)
- SpyGlass Lens ViewPort (1 page)
- MIKRON M7815 (6 pages)
- MIKRON M7816 (6 pages)
- BoilerSpection-IM Manual (43 pages)
- BoilerSpection-IM Quick Start Guide (4 pages)
- BoilerSpection-HD (90 pages)
- M390S (32 pages)
- MIKRON M345X12 (18 pages)
- M390 (35 pages)
- M330 (45 pages)
- M335 (19 pages)
- M360 (21 pages)
- M300 (19 pages)
- M310 (35 pages)
- M340 (30 pages)
- INNOVA 1316A-1 (98 pages)
- INNOVA 1316A-2 (97 pages)
- INNOVA 1316A-3 (96 pages)
- INNOVA 1303 (60 pages)
- INNOVA 1412i (202 pages)
- INNOVA 1309 (42 pages)
- INNOVA 1314i (199 pages)