4 basic settings – LumaSense Technologies IN 6/78-L User Manual
Page 22
IN 6/78-L Operating Manual
Software InfraWin · 22
5.4 Basic settings
All preset values for the
device can be displayed
and modified, if necessary under
the Devices/Parameters window.
This window contains the
parameter settings described in
Chapter 4, Parameters.
Changing an existing pyrometer
setting can be accomplished by
typing a value in an input box or
by selecting a preset value from
the list field. Choose the correct
settings for your application
from the displayed options.
Once a value is changed, it is
transferred to the pyrometer
immediately.
Clicking on the Test icon will
open a window that allows you
to directly communicate with the
pyrometer using the UPP
(Universal Pyrometer Protocol)
Data Format commands. See
Chapter 7 for information on
working with the UPP Data
Format commands.
After entering an interface
command (00 is the adjusted
address ex works, ms is the
command reading
temperature value) and
clicking on Send, the window
shown to the right will be
opened.
This window already shows the answer of the pyrometer in 1/10°. The actual temperature
reading is 399.0 °C.
Len indicates the length of the answered data string, incl. Carriage Return (Chr(13)).
In the lower part of the window the connection with the preset baud rate can be checked. Here
the command was sent 100 times with 19200 baud. It has taken 0.937 seconds without
transmission errors.