Use the Pressure Standard to output 0.00 psi. After the output has

stabilized, press the ENTER key to continue. As the unit takes

readings, the screen will show ———-. When the readings are

complete the screen should look as shown below.

Note: Only some ranges use vacuum calibration. If your gauge is not

one, than this step will be automatically skipped and calibration will be

complete.
Use the Pressure Standard to output -12.00 psi. After the output has

stabilized, press the ENTER key to continue. As the unit takes

readings, the screen will show ———-. When the readings are

complete the unit will reset and power up in normal mode.

PSI

-12.00

CAL MODE

PSI

0.00

CAL MODE