Approvals

Class I, Div. 2, Groups A-D CE approved

Appendix 1: CONDEC DPDG Calibration Procedure

Overview
Calibration adjustment of the CONDEC DPDG is performed

electronically via internal software with the case closed. There are no

mechanical adjustments; all calibration commands and adjustments

are done via the keypad, using the display to guide the user through

the calibration process.
Eight calibration points are used in the adjustment program, working

from full scale to zero at pressures equaling 100%, 87.5%, 75%, 62.5%,

50%, 37.5%, 25%, 12.5%, and 0% of full scale plus vacuum.

Note: This is an ambient temperature calibration, and should be

performed at an ambient temperature of 23° C ± 3° C (72° F ± 5° F).

Calibration outside this temperature range will invalidate the

temperature compensation program in the CONDEC DPDG.

Calibration Interval
You should check performance of the DPDG at the interval required by

your calibration program. We recommend adjustment when

measurement deviates by more than 75% of the specified accuracy, or

0.04%

Test Equipment
Verification and calibration of the CONDEC DPDG requires pressure

and/or vacuum standards able to produce and indicate pressures from

vacuum to the full-scale range of the unit under test. In order to

maintain the specified accuracy of the CONDEC DPDG, standards

should have a TUR of 4:1 or better.

Connections:
The DPDG uses a 1/4 NPT male connection in the pressure input port.

Various adapters may or may not be needed to connect to the

pressure standard. Always make sure the hose, tubing, and fittings etc

have a rated working pressure at or above the pressure of the unit.

Also it is important that there be no leaks when performing calibration;

use Teflon tape where appropriate.

Entering Calibration Mode:
After you have made your connections, turn the power on while

holding the CONFIG key. Use the arrow keys to enter the password.

The password is 101. If you have entered calibration mode correctly

9