10

the display should look as shown below. The pressure value displayed

will be the full-scale value of the gauge.

Procedure:
Screens shown in this manual represent the displays shown with a 500

psi Gauge. The CONDEC DPDG will prompt the technician for the

appropriate pressure at each calibration point.

Use the Pressure Standard to output 500.00 psi (100%). After the

output has stabilized, press the ENTER key to continue. As the unit

takes readings, the screen will show ———-. When the readings are

complete the screen should look as shown in the illustration that

follows.

Use the Pressure Standard to output 437.50 psi (87.5%). After the

output has stabilized, press the ENTER key to continue. As the unit

takes readings, the screen will show ———-. When the readings are

complete the screen should look as shown below.

Use the Pressure Standard to output 375.00 psi (75%). After the

output has stabilized, press the ENTER key to continue. As the unit

takes readings, the screen will show ———-. When the readings are

complete the screen should look as shown below.

PSI

437.50

CAL MODE

PSI

375.00

CAL MODE

PSI

312.50

CAL MODE

PSI

500.00

CAL MODE