Yokogawa SPS24 User Manual
Page 31

<6. Setup>
6-5
IM 12A01S02-01E
1st Edition : May 10, 2013-00
Table 6.3
pH/ORP common settings (numbers in parentheses indicate related sections of FLXA21
User’s Manual (IM 12A01A02-01E))
Setting Item
Available Option
Initial Value
Remarks
pH+ORP measurement
pH, ORP, pH+ORP,
pH+rH, rH
pH+ORP
Time Axis Range(Max)
1 to 10 min.
5 min.
Timeline displayed
on the calibration
screen.
pH
calibration
settings
Zero and
Slope units
Zero unit
mV, pH
mV
Slope unit
%, mV/pH
%
Select buffer set
NIST/DIN 19266,
DIN 19267, US,
Programmable
NIST/DIN 19266
Setup
0.00 to 20.00 pH
(Table 6.4)
Appears when
selecting
“Programmable”.
Temperature difference
1.0 to 5.0 degC
-1.8 to 9.0 degF
2.0 degC
3.6 degF
Sets the maximum
temperature offset
between sensors for
multiple calibration.
Limits
Zero high
0.00 to 500.0 mV
0.00 to 8.45 pH
120.0 mV
2.03 pH
Zero low
-500.0 to 0.00 mV
-8.45 to 0.00 pH
-120.0 mV
2.03 pH
Slope high
100.0 to 110.0 %
59.16 to 65.08 mV/pH
110.0 %
65.08 mV/pH
Slope low
70.0 to 100.0 %
41.41 to 59.16 mV/pH
70.00 %
41.41 mV/pH
Step Range
0.01 to 1.00 pH
0.01 pH
Stabilization
time
2 to 30 s
10 s
ORP
calibration
settings
Limits
Zero high
0.00 to 500.0 mV
120.0 mV
Zero low
-500.0 to 0.00 mV
-120.0 mV
Slope high
100.0 to 110.0 %
110.0 %
Slope low
70.0 to 100.0 %
70.00 %
Step Range
1 to 100 mV
1 mV
Stabilization
time
2 to 30 s
10 s
rH
calibration
settings
Limits
Zero high
0.00 to 500.0 mV
120.0 mV
Zero low
-500.0 to 0.00 mV
-120.0 mV
Slope high
100.0 to 110.0 %
110.0 %
Slope low
70.0 to 100.0 %
70.00 %
Step Range
0.01 to 10.00 rH
0.05 rH
Stabilization time
2 to 30 s
10 s
Impedance
settings
Impedance 1 high limit
1.000 kOhm to
1.000 MOhm
(> Impedance 1 low
limit)
200.0 kOhm
When Impedance 1
is “low*.”
In the case of
“high,” use the
value of FINE limit
for impedance
1 in SENCOM
diagnostic criteria.
Impedance 1 low limit
1.000 kOhm to
1.000 MOhm
(< Impedance 1 high
limit)
1.000 kOhm
Impedance 2 high limit
1.000 kOhm to
1.000 MOhm
(> Impedance 2 low
limit)
200.0 kOhm
When Impedance 2
is “low*.”
In the case of
“high,” use the
value of FINE limit
for impedance
2 in SENCOM
diagnostic criteria.
Impedance 2 low limit
1.000 kOhm to
1.000 MOhm
(< Impedance 2 high
limit)
1.000 kOhm