<3. Sensor Management>

3-5

IM 12A01S02-01E

1st Edition : May 10, 2013-00

Limitations

• Three different buffer solutions whose difference in pH value between buffer solutions is 1

pH or more should be used.

(1st buffer < 2nd buffer < 3rd buffer or 1st buffer > 2nd buffer > 3rd buffer)
• The temperature difference between the 1st and 2nd buffer solutions should be 20ºC or

less.

• The temperature difference between the 2nd and 3rd buffer solutions should be 20ºC or

less.

Table 3.4

Buffer selection for 3-point calibration

Buffer for 3-point Calibration

Buffer Sequence

NIST/DIN19266

pH1.7→pH6.9→pH9.2
pH4.0→pH6.9→pH9.2
pH9.2→pH6.9→pH1.7
pH9.2→pH6.9→pH4.0

DIN19267

pH4.7→pH6.8→pH9.2
pH9.2→pH6.8→pH4.7

US

pH4.0→pH7.0→pH10.0
pH10.0→pH7.0→pH4.0

User setting

Buffer Table 1 → Buffer Table 2 → Buffer Table 3
Buffer Table 3 → Buffer Table 2 → Buffer Table 1

*

In 2-point calibration, set the sequence for each buffer.

n

Temperature Calibration (Temperature Setting)

For the most accurate measurements, it is important to have a precise temperature

measurement. Measure the temperature with a high-precision thermometer and adjust the

sensor reading accordingly. For best accuracy, this should be done as near to the normal

operating temperature as possible.

n

ORP Calibration (rH Calibration)

No automatic calibration feature is available in an ORP or rH calibration.
Calibration is performed stepwise. Follow the prompts displayed on the screen.
A stability check is made at each measurement point. Proceed to the next step only after the

reading has stabilized.