10.2.4 Zero Pressure Switch, M1133AR, Figure 10.2.6


The pressure switch is used to alert the Operator via a contact input that the zero gas cylinder needs to be replaced before the
cylinder is empty. The recommended connection is to the high pressure side of the zero gas regulator. Refer to Figure 4.1.

Figure 10.2.6 Zero Pressure Switch, M1133AR

Adjustable Set

30 to 575 psi on fall;

Point Ranges:

50 to 600 psi on rise.

Dead Band:

8 to 60 psi

Over Range:

2,500 psi

Proof Pressure:

3,000 psi

Storage Temperature:

-40ºF to 180ºF (-40 ºC to 82 ºC)

Ambient Temperature:

0 ºF to 160 ºF (-18 ºC to 71 ºC)

Set Point Repeatability:

1% of span

Switch Output:

One SPDT (either NC or NO)

Electrical Rating:

Rated to 5 amps resistive and 15 amps inductive (75% PF), at 125 and 120 VAC, 1/4” HP

Enclosure:

Aluminum with irradiate finish rated for 100 hours of salt spray

Electrical Connection:

One SPDT output; 1/4” NPTF, 5ft cable

Pressure Connection:

1/8” NPTM

Mounting:

NPTM pressure connection

Weight:

12oz (340g)

Approvals:

UL 508 listed, E42272; CSA C22.2, No 14-1987: LR 9690

10.2.5 Fuses, M1132HG

Two fuses are imbedded in the power supply input terminals of the AC4 and AC8 calibration units.

Current Rating:

1 amp

Slow Blow Time:

100% - 4hours maximum

135% - 1 hour maximum

200% - 5 second maximum

Size:

1’4” O.D. x 1 1/4” L

Approvals:

UL Listed, CSA approved