2 3-3 <3 installation and wiring – Yokogawa Extractive Oxygen Gas Analyzer TDLS220 User Manual
Page 13
IM 11Y01B02-11E-A 3rd Edition :August 14, 2012-00
3-2
3-3
<3 INSTALLATION AND WIRING>
3.3 Wiring Details
Screw Terminal Block (SAK2.5) Details: (also see following diagram for internal wiring).
TB
Field or
Factory
Terminal
No.
Function
Notes/Comments
1
Factory
1
+24 VDC
DC power from the 24 power supply in adjacent
enclosure.
2
0 VDC
3
+24 VDC
24VDC power take-off to integral purge indicator/switch unit
(when fitted)
4
0 VDC
2
Field
1
SV 1 remote
Cal/Val initiate signal loop (SV # 1) to Remote voltage free con-
tacts/switch. Do not apply external power!
2
3
SV 2 remote
Cal/Val initiate signal loop (SV # 2) to Remote voltage free con-
tacts/switch. Do not apply external power!
4
5
SV 3 remote
Cal/Val initiate signal loop (SV # 2) to Remote voltage free con-
tacts/switch. Do not apply external power!
6
3
Field
1
SV 1 +
24VDC power output signal to actuate external solenoid valve # 1.
Max 12 Watts requires ferrite coil on wires.
2
SV 1 -
3
SV 2 +
24VDC power output signal to actuate external solenoid valve # 2.
Max 12 Watts requires ferrite coil on wires.
4
SV 2 -
5
SV 3 +
24VDC power output signal to actuate external solenoid valve # 3.
Max 12 Watts requires ferrite coil on wires.
6
SV 3 -
4
Field
1
Warning
Alarm
NC - Closed contact on Warning/Power-off state
2
C - Common contact – rated max 1A @ 24 VDC
3
NO - Open contact on Warning/Power-off state
4
Fault
Alarm
NC - Closed contact on Warning/Power-off state
5
C - Common contact – rated max 1A @ 24 VDC
6
NO - Open contact on Warning/Power-off state
5
Field
1
User Alarm
NC - Closed contact on Warning/Power-off state
2
C - Common contact – rated max 1A @ 24 VDC
3
NO - Open contact on Warning/Power-off state
4
Purge Alarm
C - Common contact – rated max 265 V AC/DC, 150 mA
5
NO - Open contact on loss of purge pressure