Yokogawa MLX User Manual
Page 40
Page 40 of 49
IM 60A02S01-01E-A
The numeric display shows the current engineering unit zero value with the least
significant digit blinking. At this point, the menu buttons are redefined as in Table 5.2.
Table 5.2
MLX key functions (when in Engineering units mode)
Key
Function
Selects the digit to modify (digits advance from least significant to most
significant to the sign symbol and back to least significant – 6 digits plus
sign).
Increments the selected digit (values roll over from 9 to 0).
Decrements the selected digit (values roll over from 0 to 9).
Accepts the modified value as the current engineering unit zero or full-scale
(FS) value.
(k) Touch
to increment the least significant digit to the desired valued (the digits roll over
from 9 to 0).
(l) Touch
to move to the next digit.
(m) Touch
to increment the selected (blinking) digit to the desired valued (or Touch
to decrement the selected digit).
(n) Repeat Steps (l) and (m) until desired value is indicated (including sign value) In this
example, set to 200.
(o) Touch
to accept the entered value as the engineering units zero value.
The alphanumeric display shows “Eng units 0 value”. The numeric display shows the
current engineering unit zero value.
(p) Touch
to move to the next menu item.
The alphanumeric display reads “Eng units FS value”.
(q) Touch
to set the Full Scale (FS) value.
The numeric display shows the current engineering unit FS value with the least significant
digit blinking.
(r) Repeat Steps (k) through (o) set the engineering units FS value (in this example, set to
400).
(s) Touch
to end the engineering units setting process.
The alphanumeric display reads “Return”.
(t) Touch
to exit the setup routine, save the new values and resume measurements.
The alphanumeric display shows the units selected during set up. The numeric display
resumes displaying measured values.