3 setting parameters, 1) tag no. setup, 2) calibration range setup – Yokogawa EJA118 User Manual

Page 36: Example: set a tag no. to fic-1a, Example: change the unit from mmh, O to kpa

background image

IM 1C22H1-01E

7-7

7. BRAIN TERMINAL BT200 OPERATION

7.3.3

Setting Parameters

Set or change the parameters as necessary. After
completing these, do not fail to use the “DIAG” key to
confirm that “GOOD” is displayed for the self-
diagnostic result at _

60: SELF CHECK

.

(1) Tag No. Setup (C10: TAG NO)

Use the procedure below to change the Tag No. Up to
16 alphanumeric characters can be entered.

Press the key to turn on
the BT200.

Connect DPharp and BT200 using
a communication cable and press
the key.

PARAM
01:MODEL
EJA118-DM
02:TAG NO.
YOKOGAWA
03:SELF CHECK
GOOD

OK

MENU
A:DISPLAY
B:SENSOR TYPE

HOME

SET

ADJ

ESC

MENU
C:SETTING
D:AUX SET 1
E:AUX SET 2
H:AUTO SET

HOME

SET

ADJ

ESC

MENU
C10:TAG NO.
YOKOGAWA
C20:PRESS UNIT
kPa
C21:LOW RANGE
0 kPa

DATA

DIAG

PRNT

ESC

Select C: SETTING and press the
key.

Select C10: TAG NO. and press
the key.

Displays the name of connected
DPharp model, TAG NO. and
diagnostics information. Press the
(OK) key after confirmation.

Press the (SET) key to
display the SET menu panel.

––WELCOME––

BRAIN TERMINAL
ID: BT200

check connection
push ENTER key

UTIL

FEED

F0710.EPS

SET
C10:TAG NO.
YOKOGAWA
YOKOGAWA

CODE

CAPS

CLR

ESC

When you have made an entry mistake, return
the cursor using the key, then reenter.

Set the new TAG NO. (FIC-1a).

FOKOGAWA

FIKOGAWA

FICOGAWA

FIC-GAWA

FIC-1AWA

FIC-1aWA

FIC-1a

Set TAG NO. and press the
key.

SET
C10:TAG NO.
YOKOGAWA
FIC-1a _

CODE

caps

CLE

ESC

• Example: Set a Tag No. to FIC-1a

SET
C10:TAG NO.
FIC-1a

FEED

NO

OK

DATA

DIAG

PRNT

ESC

This is the panel for confirming

set data. The set data items flash.

When all items have been confir-

med, press the

again. (To go back to the setting

panel, press the (NO) key.

SET
C10:TAG NO.
YOKOGAWA
FIC-1a

PRINTER OFF
F2:PRINTER ON

PARAM
C10:TAG NO.
FIC-1a
C20:PRESS UNIT
kPa
C21:LOW RANGE
0 kPa

FEED

POFF

NO

The DPharp TAG NO. was

overwritten.

Press the (OK) key to

return to the parameter panel.

Press the (NO) key to

return to the setting panel.

F0711.EPS

(2) Calibration Range Setup

a. Setting Calibration Range Unit

(C20: PRESS UNIT)

The unit is set at the factory before shipment if
specified at the time of order. Follow the procedure
below to change the unit.

mmH

2

O

mmAq

mmWG

mmHg

Torr

kPa

MPa

mbar

bar

gf/cm

2

kgf/cm

2

inH

2

O

inHg

ftH

2

O

psi

atm

Pa

hPa

• Example: Change the unit from mmH

2

O to kPa.

ESC

SET
C20:PRESS UNIT
kPa

FEED

NO

OK

SET
C20:PRESS UNIT
mmH20
< mmWG >
< mmHG >
< Torr >
< kPa >

Use the or

key to select “kPa.”

Press the key twice

to enter the setting.

Press the (OK) key.

F0712.EPS