Calibration menu, Thermocouple field calibration procedure – Watlow Series V4 User Manual

Page 14

background image

Calibration Menu

Figure 14 -
Calibration Parameters.

NOTE:
Before calibration on an
installed controller,
make sure all data and
parameters are docu-
mented.

Equipment Required

• Type "J" Reference Compensator with reference junction at 32°F/0°C, or Type "J"

Thermocouple Calibrator set at 32°F/0°C.

• Precision millivolt source, 0-50mV min. range, 0.01mV resolution

Setup And Calibration

1. Connect the ac line voltage L1 and L2 to the proper terminals.

2. Connect the millivolt source to Terminal 6 Negative and Terminal 5 Positive on the

Series V4 terminal block. Use regular 20 - 24 gauge wire. Make sure the DIP switch is
set for thermocouple input.

3. Apply power to the controller and allow it to warm up for 15 minutes. After warm-up

put the controller in the Calibration Menu. Select [`YES].

4. Press the ˆInfinity key twice to enter the manual mode. The controller is calibrating

when % indicator light is on. Make sure the controller is in manual mode only when
you are in the correct parameters.

5. At the 0.00 prompt, enter 0.00mV from the millivolt source to the control. Allow at

least 10 seconds to stabilize. Press the

Advance key.

6. At the 50.0 prompt, enter 50.00mV from the millivolt source to the Series V4. Allow at

least 10 seconds to stabilize. Press the ‰Advance key.

7. At the [``tC] prompt, disconnect the millivolt source, and connect the reference com-

pensator or thermocouple calibrator to Terminal 6 Negative, and Terminal 5 Positive
on the Series V4 terminal block. If using a compensator, turn on and short the input
wires. If using "J" calibrator, set to simulate 32°F/0°C. Allow 10 seconds for the control
to stabilize. The controller will leave the [`CAL] mode if one minute passes between
key activations. Press the ˆInfinity key twice to exit the manual mode. To conclude
the thermocouple calibration, advance the

Advance key to the next prompt or exit

the [`CAL] menu.

Thermocouple Field Calibration Procedure