Calibrating the series sd, Thermocouple input procedure, Rtd input procedure – Watlow Series SD User Manual
Page 6: Voltage process input procedure, Current process input procedure
Wa t l o w S e r i e s S D C a l i b r a t i o n
■
6
■
F a c t o r y a n d C a l i b r a t i o n
Calibrating the Series SD
Warm up the unit for 20 minutes. To reach the cali-
bration prompts, enter the Factory Page by holding
down the Up Key
¿ and Down Key ¯ for six seconds.
Once in the Factory Page
[FACT], use the Advance ‰
key to select a prompt. The last prompts on the Factory
Page are the input and output calibration prompts.
You can restore the original factory calibration with
Restore Factory Calibration
[rESt] (Factory Page).
*Note: INFOSENSE
TM
should be turned off while verifying calibration
of the controller with a calibration source.
Thermocouple Input Procedure
Equipment
•
Type J reference compensator with reference junc-
tion at 0°C (32°F), or type J thermocouple calibra-
tor to 0°C (32°F).
•
Precision millivolt source, 0 to 50 mV minimum
range, 0.002 mV resolution.
Input Setup and Calibration
1.
Connect the correct power supply to terminals 1
and 2 (see Chapter Two).
2.
Connect the millivolt source to terminals 11 (-) and
10 (+) with copper wire.
3.
Enter 50.00 mV from the millivolt source. Allow at
least 10 seconds to stabilize. Set Thermocouple Cal-
ibration, 50 mV
[tc;50] to [`yES]. Press the Ad-
vance Key
‰ to store 50.000 mV input and move to
the next prompt.
4.
Enter 0.000 mV from the millivolt source. Allow at
least 10 seconds to stabilize. Set Thermocouple Cal-
ibration, 0 mV
[tc;00] to [`yES]. Press the Ad-
vance Key
‰ to store 0.000 mV input and move to
the next prompt.
5.
Disconnect the millivolt source and connect the ref-
erence compensator or thermocouple calibrator to
terminals 11 (-) and 10 (+). With type J thermocou-
ple wire, if using a compensator, turn it on and
short the input wires. When using a type J calibra-
tor, set it to simulate 0°C (32°F). Allow 10 seconds
for the controller to stabilize. Set Thermocouple
Calibration, 32°
[tc;32] to [`yES]. Press the Ad-
vance Key
‰ to store type J thermocouple calibra-
tion and move to the next prompt.
6.
Rewire for operation and verify calibration.
RTD Input Procedure
Equipment
•
1 k
Ω
decade box with 0.01
Ω
resolution.
Input Setup and Calibration
1.
Connect the correct power supply to terminals 1
and 2 (see Chapter 2).
2.
Connect the decade box to terminals 10 (S1), 11
(S3) and 8 (S2), with 20 to 24-gauge wire.
3.
Enter 15.00
Ω
from the decade box. Allow at least
10 seconds to stabilize. Set RTD Calibration, 15
Ω
[`r;15] to [`yES]. Press the Advance Key ‰ to
store the 15.00
Ω
input and move to the next
prompt.
4.
Enter 380.00
Ω
from the decade box. Allow at least
10 seconds to stabilize. Set RTD Calibration, 380
Ω
[r;380] to [`yES]. Press the Advance Key ‰ to
store the 380.00
Ω
input and move to the next
prompt.
5.
Rewire for operation and verify calibration.
Voltage Process Input Procedure
Equipment
•
Precision voltage source, 0 to 10V minimum range,
with 0.001V resolution.
Input Setup and Calibration
1.
Connect the correct power supply to terminals 1
and 2 (see Chapter Two).
Input
2.
Connect the voltage source to terminals 11 (-) and
9 (+) of the controller.
3.
Enter 1.00V from the voltage source to the con-
troller. Allow at least 10 seconds to stabilize. Set In-
put Calibrate, 1V
[``u;1] to [`yES]. Press the Ad-
vance Key
‰ to store the 1.00V input and move to
the next prompt.
4.
Enter 9.00V from the voltage source to the con-
troller. Allow at least 10 seconds to stabilize. Set In-
put Calibrate, 9V
[``u;9] to [`yES]. Press the Ad-
vance Key
‰ to store the 9.00V input and move to
the next prompt.
5.
Rewire for operation and verify calibration.
Current Process Input Procedure
Equipment
•
Precision current source, 0 to 20 mA range, with
0.01 mA resolution.
Input Setup and Calibration
1.
Connect the correct power supply to terminals 1
and 2 (see Chapter Two).
2.
Connect the current source to terminals 11(-) and
8(+).
3.
Enter 4.00 mA from the current source to the con-
troller. Allow at least 10 seconds to stabilize. Set In-
put Calibrate, 4 mA
[``A;4] to [`yES]. Press the
Advance Key
‰ to store the 4 mA input and move
to the next prompt.
4.
Enter 16.00 mA from the current source to the con-
troller. Allow at least 10 seconds to stabilize. Set In-
put Calibrate, 16 mA
[`A;16] to [`yES]. Press the
Advance Key
‰ to store the 16 mA input and move
to the next prompt.
5.
Rewire for operation and verify calibration.