ZOLL R Series Monitor Defibrillator Rev D User Manual
Page 19

Specifications
9650-0901-01 Rev. D
R Series Pulse Oximetry (SpO
2
)
17
Note:
The Pulse Oximetry Option is calibrated for functional saturation.
Note:
Because R Series pulse oximeter measurements are statistically distributed, only about
68% of these measurements can be expected to fall within plus or minus one standard
deviation of the value measured by a CO-oximeter.
* The accuracy of the R Series pulse oximeter can only be verified by comparing its
measurements to those obtained with a CO-oximeter. Pulse oximeter functional testers cannot
be used to verify the accuracy of the oximeter or its sensors.
Pulse rate accuracy
During no-motion conditions
25 to 240 beats per minute, ±3 beats per minute
During motion conditions
25 to 240 beats per minute, ±5 beats per minute
Pulse rate alarm limits
On/Off displayed on monitor; operator-selectable.
Low limit:
20 to 100 beats per minute
High limit:
60 to 235 beats per minute
Biocompatibility
Patient-contacting material meets requirements of
ISO 10993-1, Biological Evaluation of Medical Device,
Part I, for external surfaces, intact surfaces, and
short-term exposure.
Environmental
Temperature
Operating:
0°C to 40°C
Storage and shipping:
–20°C to 60°C
Note: The defibrillator might not perform to
specifications if stored at the upper or lower
temperature extreme and then put to immediate
use.
Electromagnetic immunity
EN60601-2-4; IEC 1000-4-3 to 18 V/m
Software Hazards
Minimized by compliance with EN14971