Define format used to set trace names

Stores a saved waveform naming format or convention The choices are LeCroy or Dialog. The LeCroy
choice produces waveform names automatically by the software. Dialog prompts the user for custom
waveform names. The default setting is LeCroy.

Use Stored Trace for Speed Grade

This is an optimization used specifically to measure the clock frequency only once. Choose from Yes or
No values, The default selection is No.

Advanced Settings

Clock Period per Screen Division

Oscilloscope timebase and sampling rate is set to acquire the given number of clock cycle per display
horizontal division at a given DUT Speed Grad in MT/s and for a Max. Number of Samples Per Clock
Period
. The default is 3341 clock periods (this gives a 10us/div timebase at 667 MT/s and 3.3MS max for
100 samples per period).

Timebase = [Clock Period Per Screen Division] / ([DUT Speed Grade in MT/s] / 2 * 1e6)

Maximum Samples = [Max. Number Of Samples Per Clock Period] * [Clock Period Per Screen
Division]
* 10

Number of cycles for Clock test

Jedec standard requires 200 cycles for the Clock compliance test.

This is the default value of this variable. Any positive number can be entered.

Max. Number Of Samples Per Clock Period

The oscilloscope timebase and sampling rate is set to acquire the given number of points per clock
period. The oscilloscope is always set to at least acquire at 20GS/s. Additionally, if an oscilloscope with
greater than 6GHz bandwidth is used, the bandwidth is limited to 6GHz. See the Clock Period Per
Screen Division
topic for more details. Choose between 10;20;50;100;200;500 or 1000. The default
value is 100.

Configuration Specific Variables

The following variables are specific to the configuration in which they appear under. Some of these
variables appear under multiple configurations.

XX Channel Gain

Allows the user to manually specify the vertical scale in V/div for XX SUT. XX can be Clock, DQ, DQS,
DQSn, ADD/CTRL, or DM. Default is 0 for auto-scale.

XX Channel Index

Allows the user to manually specify the channel XX SUT. XX can be Clock, DQ, DQS, DQSn, ADD/CTRL,
or DM. Default is 1 for CK, 2 for DQS, 3 for DQ and 4 for others

XX Channel Invert

Allows the user to invert XX SUT. XX can be Clock, DQ, DQS, DQSn, ADD/CTRL, or DM. Default is
False.

XX Channel Offset

Allows the user to manually specify the offset in Volts for XX SUT. XX can be Clock, DQ, DQS, DQSn,
ADD/CTRL, or DM. Default is 0 for auto-scale.