Operating environment – Teledyne LeCroy CP500 User Manual
Page 8

CP500 Current Probe
2
922176-00 Rev A
Never install or remove the probe on bare conductors which are energized. The transformer
core and shield are grounded but not insulated and may contact the conductor when the locking
lever is open.
To avoid short circuits and accidents that could result in injury or death, use the CP500 current
probe only with conductors carrying 600 V (CAT II) or 300 V (CAT III) or less.
To prevent short circuits and electric shock when conductors being measured carry less than
600 V (CAT II) or 300 V (CAT III), but more than the safe voltage level (SELV-E) make sure that the
conductors to be measured are insulated with material conforming to:
o Overvoltage Category III, basic insulation requirement for working voltage of 300 V
o Pollution degree 2
Operating Environment
The accessory is intended for indoor use and should be operated in a clean, dry environment. Before
using this product, ensure that its operating environment is maintained within these parameters:
Temperature: 0° to 40° C
Humidity: ≤ 80% relative humidity (non-condensing)
Altitude: up to 2000 m (6562 feet)