Typical characteristics, Environmental characteristics – Teledyne LeCroy CP500 User Manual
Page 23

Instruction Manual
922176-00 Rev A
17
Typical Characteristics
Typical characteristics describe parameters which do not have guaranteed performance; however are
representative of the average performance from a sample of several probes. Tests for typical
characteristics are not provided in the Performance Verification procedure.
Sensitivity*
200 mA/div* to 200 A/div.
Scale Factor Temperature
±2% or less. (from 0° to +40 °C,
Coefficient
32 °F to 104 °F)
Noise
Equivalent to 25 mA
rms
or less
(Bandwidth of measuring instrument:
20 MHz)
Insertion Impedence
Refer to Figure 6
Bandwidth (-3 dB)
DC to 2 MHz
Rise Time
≤ 175 ns
* Lower sensitivity may change with different oscilloscope models.
Environmental Characteristics
Operating Temperature Humidity
0 to 40 °C (32 °F to 104 °F)
≤ 80% relative humidity (non-condensing)
Storage Temperature Humidity
–10 °C to 50 °C (14 °F to 122 °F)
≤ 80% relative humidity (non-condensing)
Usage
Indoor
Altitude
up to 2000 m (6562 feet)
Effect of External Magnetic
Equivalent to a maximum of
Field
800 mA (in a DC or 60 Hz,
400 A/m magnetic field)