CP500 Current Probe

14

922176-00 Rev A

Table 2. List of Adjustment Procedure equipment.

Description

Minimum Requirements

Test Equipment Examples

Wide Band Oscilloscope

Minimum 200 MHz bandwidth

ProBus interface equipped

Software version 4.3.0.0 or higher

Teledyne LeCroy WaveRunner

6030A

Digital Multimeter (2 required)

DC: 0.1% Accuracy

5½ digit resolution

Agilent Technologies 34401A, or

Fluke 8842A-09

Function Generator

50 Hz sine wave output
3 Vrms into 50 Ω

Agilent Technologies 33120A, or

Stanford Research Model DS340

Calibration Fixture, 100 Turn Loop

100 Turn loop in series with 0.5 Ω
±0.1% resistor with sense terminals

Teledyne LeCroy CP500-CF02

Banana Plug Adapter

Female BNC to Dual Banana Plug

Pomona 1269

Calibration Fixture

ProBus Extension Cable

Teledyne LeCroy PROBUS-CF01

Patch Cables (4 required)

Male Banana to Male Banana, 12"

Pomona B-12-0 (black) or

Pomona B-12-2 (red)

BNC Adapter

BNC Male to Dual Banana Jack

Pomona 1296

Preliminary Procedure

1. Remove the probe compensation box circuit board by removing the two screws from the cable

end of the compensation box and sliding the circuit board out of the box.

2. Connect the CP500 compensation board to the channel 1 input of the oscilloscope through the

ProBUS extension cable, and close and lock the probe jaws.

3. Turn the oscilloscope on and allow at least 30 minutes warmup time for the CP500 and test

equipment before performing the Verification Procedure.