Certifications
This section contains the instrument’s Electromagnetic Compatibility (EMC),
Safety and Environmental certifications.
EMC Compliance
EC Declaration of Conformity - EMC
The probe meets intent of EC Directive 2004/108/EC for Electromagnetic
Compatibility. Compliance was demonstrated to the following specifications as
listed in the Official Journal of the European Communities:
EN 61326-1:2006, EN 61326-2-1:2006 EMC requirements for electrical
equipment for measurement, control, and laboratory use.
Electromagnetic Emissions:
CISPR 11:2003, Radiated and Conducted Emissions Group 1, Class A
1 2
Electromagnetic Immunity:
EN 61000-4-2:2001 Electrostatic Discharge, 4 kV contact, 8 kV air, 4 kV
vertical/horizontal coupling planes
3
EN 61000-4-3:2006 RF Radiated Electromagnetic Field, 3 V/m, 80-1000 MHz; 3
V/m, 1400 MHz - 2 GHz; 1 V/m, 2 GHz - 2.7 GHz
3
1 Emissions which exceed the levels required by this standard may occur when the probe is
connected to a test object.
2 This product is intended for use in nonresidential areas only. Use in residential areas may cause
electromagnetic interference.
3 Meets Performance Criteria “B” limits of the respective standard: during the disturbance,
product undergoes a temporary degradation or loss of function or performance which is self-
recoverable.