Teledyne LeCroy DXC100A User Manual

Page 18

background image

DXC100A Differential Probe Pair

16

922259-00 Rev A

When the +probe has been properly adjusted it should be used as a reference.
CMRR adjustments made to the probe in the future should be done to the –
probe.

At this point, do not change any of the +Input X10 or +Input X100 Adjustments,
as doing so may require repeating the entire probe calibration procedure.

Figure 1, The DXC100A Board Layout