beautypg.com

Install the lm10zi-acqmod rack mount – Teledyne LeCroy LabMaster 10Zi Rackmount User Manual

Page 6

background image

LabMaster 10Zi Rack Mount

Install the LM10Zi-ACQMOD Rack Mount

1. Use the empty screw holes in the two brackets at the left and right front sides of the tray to install it into a rack following

manufacturer specifications.

Four fasteners per side is required. The center hole in the bracket accommodates an alignment pin on the rack.

NOTE: Acquisition Modules should be mounted directly above or below the MCMZi in the rack so that the cables can reach their
intended connectors.

2. Slide the Acquisition Module into the mounted tray until it touches the two positioning feet at the back of the tray. Do not try to

push the unit over the feet.

CAUTION. Due to the weight of the LM10Zi Acquisition Module, transporting the unit while mounted in a rack is not
recommended. The instrument should always be removed from the rack and packed for transit in the required
Teledyne LeCroy shipping materials.

6

923719-00 Rev A