1. Dust Fraction correction on the SPLIT2

Extended Options
Sample Rate
Erase Memory
Cancel

Run
Review Data
Special Functions
Auto-Zero

Size Select
Battery Status
Cancel

Select
Apply Scale
Restore Default
Cancel

Respirable
Thoracic
Inhalable

System Options
Date/Time
Set Alarm
Cancel

Scroll arrow to Systems Options option and press the ENTER
button

Before a SCALE factor can be entered, a gravimetric result
must be obtained from the filter contained behind the
sensing head. The filter must be preweighed.

Scroll arrow to Extended Options and press the ENTER button

Scroll arrow to Size Select Option and press ENTER

Scroll arrow to apply scale Option and press ENTER

Scroll arrow to Respirable, Thoracic or Inhalable Option and press
ENTER. The fraction selected depends on the fraction you have
sampled and obtained a gravimetric result for.
NOTE: To do this for Respirable you must have a gravimetric
result, obtained by using the IOM with foam and a filter.

* SCALE *

Scale: 01.00

Go to the Special Functions option by pressing either the down
button or the up button until the arrow is opposite Special
Functions. Press ENTER button

A default scale of 01.00 will be seen. Enter the SCALE factor
required by using the UP, DOWN and ENTER buttons.
The screen will now return to the APPLY SCALE screen.

DUST FRACTION CORRECTION