SKC 575-005 Series Passive Sampler User Manual
Page 2
2
Sampling
1. Remove sampler from sealed pouch. Retain
and set aside all other parts, including the
desorption effi
ciency tube(s).
2. Write the start time, date, and a user-assigned
sampler ID number on the label on the back of
the sampler.
3. Remove the lock, cap, and O-ring from the
sampler and set aside. Clip sampler to worker’s
clothing in the breathing zone. Ensure small
holes are facing out.
Note: Sample the entire workshift to obtain a true 8-hour exposure. If breaks/lunch
are part of the 8-hour workday or occur in the work area, samplers may remain
sampling. If desired, the sampler can be removed for breaks/lunch and sealed with
the provided cap. When the worker returns to the work area, replace the sampler on
the worker’s clothing and remove the cap. Note the times the sampler is removed
from and placed back on the worker to ensure accurate air volumes.
4. At the end of the desired sampling period,
unclip the sampler from worker clothing.
Push the O-ring, cap, and lock (removed
during Step 3) onto the side of the sampler
containing the holes.
5. Write the stop time on the label on the back of
the sampler.
6. Carefully package the sampler and all other
parts and send to an AIHA-accredited
laboratory for analysis. If using an in-house
laboratory, refer to the analysis instructions
included in this package.
1
2
3
4
5
6