beautypg.com

SKC 590-300 Nitrous Oxide Passive Sampler User Manual

Introduction, Performance profi le

background image

Nitrous Oxide Passive Sampler

Cat. No. 590-300

Operating Instructions

863 Valley View Road, Eighty Four, PA 15330 USA

Tel: 724-941-9701 Fax: 724-941-1369 e-mail: [email protected]

Introduction

The SKC Nitrous Oxide Passive Sampler is a small sampler that provides accurate
and reliable sampling of nitrous oxide in the 50 to 450 ppm range (12.75 to 890 ppm-
hours dose range). The samplers are packaged as empty sampler housings that the
user fi lls with cleaned/purged Molecular Sieve 5A sorbent supplied in glass vials
with PTFE-lined caps. This method of packaging maintains low sorbent background.
Samples are thermally desorbed and analyzed by gas chromatography with electron
capture detector.

Performance Profi le

Sorbent/Amount:

Molecular Sieve 5A 60/80, 736 mg (supplied in glass vial)

Housing Material:

Nylon, non-sterile

Diameter:

1.4 in (3.5 cm)

Length
(including clip):

2.5 in (6.3 cm)

Depth:

0.6 in (1.5 cm)

Concentration Range: 50 to 450 ppm (12.75 to 890 ppm-hrs dose)
Analysis:

Thermal desorption; gas chromatography (GC) with
electron capture detector (ECD)

Shelf-life:

6 months from date of manufacture. Store at < 39.2 F (4 C).

Sample Storage:

3 weeks at < 39.2 F (4 C) or 1 week at ambient temperature

Sample Time:

Validated for 8-hour sampling

Sampling Rate:

0.81 ml/min ± 12% RSD

Samplers are non-sterile.

www.skcinc.com

Collar Clip

Primary

Diffusion

Barrier

Front View
of Housing

Green Secondary
Diffusion Barrier

Cover

Collar

Clip

Vial

Adapter

Back View

of Housing

Plug

Vial containing

purged sorbent