Performance profi le, Sampling instructions – SKC 500-400 UMEx 400 Passive Sampler for Amines User Manual
Page 2
Performance Profi le
Sampling Rates:
Compound
Sampling Rate
RSD
(ml/min)
Methyl
amine
18.4
14.0
Dimethyl
amine
18.2
5.8
Isopropyl
amine
13.9
9.5
Allyl
amine
22.4
7.7
n-Butyl
amine
18.1
9.2
Concentration Range: 0.5 to 10 ppm
Detection Principle:
Formation of a stable amine derivative, made from a
reaction of the amine with NITC
Analysis Method:
Analysis by HPLC with UV detection at 254 nm
Accuracy:
± 25% for all amines except methyl amine
± 35% for methyl amine
Shelf-life:
12 months from date of manufacture in freezer
Storage:
Before use: ≤39.2 F (4 C)
After use: Samples can store at ambient temperatures
for 14 days
For storage greater than 14 days, store at ≤39.2 F (4 C).
Do not store with food.
Temperature Effects: No effect on sampling rate between 50 and 86 F
(10 and 30 C)
Humidity Effects:
No effect on sampling rate from 10 to 80% relative humid-
ity (RH)
Wind Velocity Effects: No effect from 5 to 100 cm/sec
Interferences:
None found; highly specifi c for amines
Validation: Partial
Dimensions:
3.4 x 1.1 x 0.35 in (8.6 x 2.8 x .89 cm)
Weight:
0.38 oz (10.8 g)
Slide Cover: Blue
Sampling Instructions
Cautions: • Store at ≤ 39.2 F (4 C) before use.
• Do not store with food.
• Before sampling, check the expiration date on the label on the
outside of the pouch. Do not use aft er the last day of the month
indicated.
• SKC recommends using gloves when handling chemically
treated media.
•
UME
x
samplers are designed for single use. Do NOT reuse
UME
x
samplers.