beautypg.com

Performance profi le, Sampling instructions – SKC 500-400 UMEx 400 Passive Sampler for Amines User Manual

Page 2

background image

Performance Profi le

Sampling Rates:

Compound

Sampling Rate

RSD

(ml/min)

Methyl

amine

18.4

14.0

Dimethyl

amine

18.2

5.8

Isopropyl

amine

13.9

9.5

Allyl

amine

22.4

7.7

n-Butyl

amine

18.1

9.2

Concentration Range: 0.5 to 10 ppm

Detection Principle:

Formation of a stable amine derivative, made from a
reaction of the amine with NITC

Analysis Method:

Analysis by HPLC with UV detection at 254 nm

Accuracy:

± 25% for all amines except methyl amine

± 35% for methyl amine

Shelf-life:

12 months from date of manufacture in freezer

Storage:

Before use: ≤39.2 F (4 C)

After use: Samples can store at ambient temperatures
for 14 days

For storage greater than 14 days, store at ≤39.2 F (4 C).

Do not store with food.

Temperature Effects: No effect on sampling rate between 50 and 86 F

(10 and 30 C)

Humidity Effects:

No effect on sampling rate from 10 to 80% relative humid-
ity (RH)

Wind Velocity Effects: No effect from 5 to 100 cm/sec

Interferences:

None found; highly specifi c for amines

Validation: Partial

Dimensions:

3.4 x 1.1 x 0.35 in (8.6 x 2.8 x .89 cm)

Weight:

0.38 oz (10.8 g)

Slide Cover: Blue

Sampling Instructions

Cautions: • Store at ≤ 39.2 F (4 C) before use.

• Do not store with food.

• Before sampling, check the expiration date on the label on the

outside of the pouch. Do not use aft er the last day of the month
indicated.

• SKC recommends using gloves when handling chemically

treated media.

UME

x

samplers are designed for single use. Do NOT reuse

UME

x

samplers.