beautypg.com

Description, Performance profi le – SKC 225-9594 SKC BioSampler User Manual

Page 2

background image

Description

The patented* SKC BioSampler is a highly effi

cient bioaerosol and biologically

inert airborne particle collection device that traps airborne microorganisms into
swirling liquid for subsequent analysis. The BioSampler is made of glass and
consists of three parts: inlet, nozzle section (with three tangential sonic nozzles),
and collection vessel (Figure 1). The collection vessel can be fi lled with a liquid
collection medium or coated with a sticky medium. The BioSampler can be used
with water or non-evaporating liquids up to 1,000 times more viscous than water
for sampling up to eight hours. The BioSampler requires a high-volume vacuum
pump such as the SKC VAC-U-GO Sampler.

* U.S. Patent Nos. 5,902,385 and 5,904,752

Performance Profi le

Flow Rate:

Approximately 12.5 L/min

Medium:

Water or more viscous

liquids such as ViaTrap

®

mineral

oil

Collection
Effi

ciency:

Nearly 100% over a
wide range of particle
sizes; decreases to
approximately 90% at
0.5 μm

Pump:

High-volume vacuum
pump capable of
achieving sonic fl ow

† May not be suitable for PCR analysis

Inlet
(aerosol in)

Sonic Nozzles (3)

To pump

Figure 1. BioSampler with 20-ml vessel

(5-ml vessel available)

1