SPECIFICATION
CO

2

Range:

0-5,000ppm
5,001-9,999 (out of scale)

Resolution:

1ppm

Accuracy:

±50ppm ±5%rdg (0-5,000ppm)
Not specified for out-of-scale

Pressure
Dependence:

+1.6% reading per kPa deviation from
normal pressure, 100kPa

TEMPERATURE

Range:

14º-140°F (-10.0º to 60.0°C)

Resolution:

0.1°F (0.1°C)

Accuracy:

±0.9°F (0.6°C)

HUMIDITY

Range:

0.0 to 99.9%

Resolution:

0.1%

Accuracy:

±3% (10 to 90%) ±5% (others)

Warm up:

30 seconds

Operating:

32° to 122°F (0º to 50°C), 0 to 95%RH
(avoid condensation)

Storage:

-4° to 140°F (-20º to 60°C), 0 to 99%RH
(avoid condensation)

Power Source:

4 “AA” batteries, DC adaptor

Battery life:

24 hours (Alkaline)

CO

2

LEVELS AND GUIDELINES

NON-ENFORCED REFERENCE LEVELS

NIOSH recommendations

250-350ppm:

Normal outdoor ambient concentrations

600ppm:

Minimal air quality complaints

600-1000ppm:

Less clearly interpreted

1000ppm:

Indicates inadequate ventilation; complaints
such as headaches, fatigue, and eye/throat
irritation will be more widespread. 1000ppm
should be used as an upper limit for indoor
levels.

EPA Taiwan:

600ppm and 1000ppm