Fairbanks 3300 Reliant Series Floor Scale User Manual

Page 16

background image

10/06

16

50783 Issue 1

3.

Loosen all gland bushing nuts

4.

Check that all load cells have wire markers on the cable ends. If not, identify
load cells with wire markers or other means, then disconnect the load cell
wires from the terminal blocks. Disconnect the main interface cable wires.

5.

Remove the balance junction box/ PCB, then install the new balance junction
box/ PCB.

6.

Reconnect all load cell and main interface cable wires to the new balance
junction box/ PCB.

7.

Tighten all gland bushing nuts

8.

Recalibrate as necessary.

9.

Replace the box cover and tighten all screws. Replace the platform access
cover .

Note:

Leave the box cover OFF until all calibration adjustments are completed.