Cable Wire Color

Function

Balance Junction Box

Black

(-) Excitation

- E

Green

(+) Excitation

+ E

Yellow

Shield

SH

White

(+) Signal

+ S

Red

(-) Signal

- S

g.

Connect the main interface cable as follows:

Cable Wire Color

Function

Balance Junction Box

Black

(-) Excitation

- EXC

Green

(+) Excitation

+ EXC

Yellow

Shield

SHLD

White

(+) Signal

+ SIG

Red

(-) Signal

- SIG

*

(-) Sense

- SEN

*

(+) Sense

+ SEN

* = Used only when main interface cables exceed twenty five (25) feet.

h.

Tighten all gland bushing nuts

i.

Calibrate as necessary. See Platform Cornering Adjustments.

j.

Replace the box cover and tighten all screws. Replace the platform
access cover .

2.

Platform Cornering Adjustments

a.

Adjust the analog indicator to the platform. First, ensure all corners are

within one (1) division of each other at 25% of rated capacity. Follow the

appropriate indicator service manual for indicator calibration.

1.)

Perform a course platform calibration close to the actual weight.

2.)

Identify the platform corner numbers.