3.)

Place a concentrated weight (25% of platform capacity) on corner 1,

move it to corner 2, corner 3 and corner 4, noting the displayed

reading on each corner.

4.)

If corners require adjustment:

a.)

Place the concentrated weight on the corner displaying the

lowest weight and use the appropriate potentiometer to

change the displayed weight to read the same as the

highest reading by turning the potentiometer clockwise

(CW). Repeat this procedure while rechecking all corners

until there are no errors. Perform a zero reference check

with a unloaded platform, then repeat the corner test to

ensure all readings are the same before proceeding.

4

4

3

3

2

2

1

1

+ SIG

- SIG

- SEN

- EXC

+ SEN

- EXC

SHLD

+E -E SH -S +S

+E -E SH -S +S +E -E SH -S +S

+E -E SH -S +S

96141

Note:

As you move the weight(s) from corner to corner, do NOT zero the scale. Remember, at this

time you are only adjusting the corners to be the same, not a performing a correct calibration.

Note:
The balance junction box assembly has four (4) extended range, multi-turn potentiometers,
one for each load cell. If you have difficulty cornering the platform, turn all potentiometers to
the fully counter clockwise (CCW) position until a clicking is heard when turned. With the
weight on the lowest reading corner, turn the corresponding potentiometer clockwise (CW) to
read the same as the highest reading corner. Repeat this procedure for each corner until the
platform is properly calibrated.