A.

Errors

Symptom Cause Solution

Corner Errors

Debris under scale.

Remove debris.

Load cell failure.

Replace load cell.

Broken foot assembly.

Replace foot.

Platform not level.

Level platform.

Load cell tolerance

Adjust balance

change.

junction box per

Section 4, if

equipped.

Unstable indication

Cables damaged.

Repair cables.

Load cell failure

Replace load cell.

Indicator failure

Repair indicator

B.

Load Cell Replacement

1.

Remove power to the indicator.

2.

Remove the cover plate. Identify the load cells with wire markers or other

means and disconnect the failed load cell cable at the splice connections.

3.

Lift the platform end with a forklift or heavy pry bar using wood blocks for safety.

4.

Remove the load cell mounting bolts (use a ¾" socket), then the cell, pulling the

load cell cable through the scale while leaving a pull string/wire in the scale.

5.

Remove the foot assembly from the old load cell and install on the new load

cell. Use anti-seize on the bolt threads.