3 Operation

Model 6000

3-24 Part I

Teledyne Analytical Instruments

3.6

The Range Select Function

The

Range

function allows you to manually select the concentration range

of analysis (MANUAL), or to select automatic range switching (AUTO).

In the MANUAL screen, you are further allowed to define the high and low

(concentration) limits of each Range, and select a single, fixed range to run.

CAUTION: If this is a linearized application, the new range must

be within the limits previously programmed using the
System function, if linearization is to apply through-
out the range. Furthermore, if the limits are too small
a part (approx 10 % or less) of the originally linear-
ized range, the linearization will be compromised.

3.6.1 Manual (Select/Define Range) Screen

The Manual range-switching mode allows you to select a single, fixed

analysis range. It then allows you to redefine the upper and lower limits, for the
range.

Press

Range

key to start the Range function.

Select

range

mode: MANUAL

If above screen displays, use the

Δ∇ arrow keys to Select MANUAL, and

press

Enter.

Select range to run

—> Ш1 Ш2 Ш3 CAL<—

Use the < > keys to select the range: 01, 02, 03, or CAL. Then press

Enter

.

Gas use: SO2

Range: Ø — 10 %

Use the < > keys to toggle between the Range: low-end field and the

Range: high-end field. Use the

ΔΔΔΔΔ∇ keys to change the values of the fields.

Press

Escape

to return to the previous screen to select or define another

range.

Press

Enter

to return the to the

Analyze

function.