1-6: Part I

1 Introduction

Model 6000B

Teledyne Analytical Instruments

Power Connection

AC power source, 115VAC, 50/60 Hz

Analog Outputs

0-1 V dc concentration and 0-1 V dc
range ID. Isolated 4-20 mA dc and 4-20
mA dc range ID.

Alarm Connections

2 concentration alarms and 1 system
alarm.

RS-232 Port

Serial digital concentration signal output
and control input.

Remote Bench

Provides all electrical interconnect to the
Analysis Unit.

Remote Span/Zero

Digital inputs allow external control of
analyzer calibration.

Calibration Contact

To notify external equipment that
instrument is being calibrated and
readings are not monitoring sample.

Range ID Contacts

Four separate, dedicated, range relay
contacts. Low, Medium, High, Cal.

Network I/O

Serial digital communications for local
network access. For future expansion.
Not implemented at this printing.

Note: If you require highly accurate Auto-Cal timing, use external

Auto-Cal control where possible. The internal clock in the
Model 6000B is accurate to 2-3 %. Accordingly, internally
scheduled calibrations can vary 2-3 % per day.