3 Operation
Model 6000A
3-22 Part I
Teledyne Analytical Instruments
–OR –
Go to
Ltch and then press either ∆ two times or ∇ two times.
(Toggle it to
N and back to Y.)
3.6
The
Range
Select Function
The
Range
function allows you to manually select the concentration range
of analysis (
MANUAL), or to select automatic range switching (AUTO).
In the
MANUAL screen, you are further allowed to define the high and low
(concentration) limits of each Range, and select a single, fixed range to run.
CAUTION: If this is a linearized application, the new range must
be within the limits previously programmed using the
System function, if linearization is to apply through-
out the range. Furthermore, if the limits are too small
a part (approx 10 % or less) of the originally linear-
ized range, the linearization will be compromised.
3.6.1 Manual (Select/Define Range) Screen
The Manual range-switching mode allows you to select a single, fixed
analysis range. It then allows you to redefine the upper and lower limits, for the
range.
Press
Range
key to start the Range function.
Select range
mode: MANUAL
If above screen displays, use the
∆∇ arrow keys to Select MANUAL, and
press
Enter.
Select range to run
> Ш1 Ш2 Ш3 CAL<
Use the < > keys to select the range: 01, 02, 03, or CAL. Then press
Enter
.
Gas use: SO2
Range: Ø 10 %
Use the < > keys to toggle between the Range: low-end field and the
Range: high-end field. Use the
∆∆∆∆∆∇ keys to change the values of the fields.