Range – Teledyne 6000A - Photometric analyzer User Manual

Page 46

background image

3 Operation

Model 6000A

3-22 Part I

Teledyne Analytical Instruments

–OR –

Go to

Ltch– and then press either ∆ two times or ∇ two times.

(Toggle it to

N and back to Y.)

3.6

The

Range

Select Function

The

Range

function allows you to manually select the concentration range

of analysis (

MANUAL), or to select automatic range switching (AUTO).

In the

MANUAL screen, you are further allowed to define the high and low

(concentration) limits of each Range, and select a single, fixed range to run.

CAUTION: If this is a linearized application, the new range must

be within the limits previously programmed using the
System function, if linearization is to apply through-
out the range. Furthermore, if the limits are too small
a part (approx 10 % or less) of the originally linear-
ized range, the linearization will be compromised.

3.6.1 Manual (Select/Define Range) Screen

The Manual range-switching mode allows you to select a single, fixed

analysis range. It then allows you to redefine the upper and lower limits, for the
range.

Press

Range

key to start the Range function.

Select range

mode: MANUAL

If above screen displays, use the

∆∇ arrow keys to Select MANUAL, and

press

Enter.

Select range to run

—> Ш1 Ш2 Ш3 CAL<—

Use the < > keys to select the range: 01, 02, 03, or CAL. Then press

Enter

.

Gas use: SO2

Range: Ø — 10 %

Use the < > keys to toggle between the Range: low-end field and the

Range: high-end field. Use the

∆∆∆∆∆∇ keys to change the values of the fields.