2 Installation

Model 6000A

2-10: Part I

Teledyne Analytical Instruments

Table 2-5: Required RS-232 Options

Parameter

Setting

Baud

2400

Byte

8 bits

Parity

none

Stop Bits

1

Message Interval

2 seconds

Remote Bench and Solenoid Valves: The 6000A is a single-chassis

instrument. However, the REMOTE BENCH and SOLENOID RETURN
connector is provided on the uper right corner of the backpanel as a 12 pin
connector. The Remote Bench is wired at the factory as well as any op-
tional solenoid valves included in the system.

2.4

Testing the System

After The Control Unit and the Analysis Unit are both installed and

interconnected, and the system gas and electrical connections are complete,
the system is ready to test. Before plugging either of the units into their
respective power sources:

Check the integrity and accuracy of the gas connections. Make
sure there are no leaks.

Check the integrity and accuracy of all electrical connections.
Make sure there are no exposed conductors

Check that sample pressure typically between 0 and 30 psig,
according to the requirements of your process.

Power up the system, and test it by performing the following operation:

1. Repeat the Self-Diagnostic Test as.