Maintenance Model

3110

Teledyne

Analytical

Instruments

32

4. Note how much the reading has changed after a few minutes.

5. Reduce the flowrate to 0.5 SCFH.

6. Note how much the reading has changed.

7. The reading should change by less than 10% of the original

value