beautypg.com

Teledyne 3110 – Portable Trace/Percent Oxygen Analyzer, FM-approved configuration User Manual

Page 32

background image

Operation Model

3110

Teledyne

Analytical

Instruments

24

1. Press ENTER to stop the data logger and return to the LOG

RESET & START screen.

3.2.11 LOG TRANSMIT Screen

Data can be downloaded to a computer using the 9-pin D-

subminiature port and cable attached to a PC. The D-sub port accepts a
standard data cable with a DB-9 connector. The computer must be able
to accept data from a source with the following characteristics:

9600 baud

8 bit data

no parity

1 stop bit

no flow control

Refer to your computer manual for details on how to setup the RS-

232 communications port on your computer.

To download a data log:

1. Use the UP/DOWN keys to navigate to the LOG TRANSMIT

screen. Then press ENTER to enter the setup screen.

2. Press ENTER again to transmit the current data log.

Note: The data set will be transmitted each time you press the

ENTER key. Use the ESC key to exit the LOG TRANSMIT