beautypg.com

Teledyne 3110 – Portable Trace/Percent Oxygen Analyzer, FM-approved configuration User Manual

Page 21

background image

Portable Oxygen Analyzer

Installation

Teledyne

Analytical

Instruments

13

remove traces of trapped air. Otherwise, trapped air in the lines
(especially between the regulator and cylinder) will cause the analyzer
reading to slowly change over a period of hours.

To bleed the regulator and sample line:

1. Attach the regulator to the gas cylinder. Then attach a sample

line with a flow control/shut off valve preferably at the far
end of the sample line.

2. Open the shut off valve slightly, and then open the valve on

the gas cylinder.

3. Adjust the regulator to the desired pressure (usually 5 psi)

then close the cylinder valve.

4. Open the cylinder valve to pressurize the regulator fully then

close the cylinder valve again.

5. Open the sample flow control valve and allow the gas to

bleed down and vent to a safe area. Observe the secondary
gauge (low pressure side) on the regulator. As the low
pressure gauge starts to fall, close the sample flow control
valve.

6. Repeat steps 4 and 5 seven (7) times.

The sample delivery system is now purged and ready for calibration

or analysis. Keep the cylinder valve open to maintain system
pressurization.

Note: Make sure there are no leaks in the sample line and

regulator connections. Check also the cylinder connection
for leaks.

2.5 Set the Sample Flowrate

Once the system has been calibrated, the instrument can be brought

to the analysis site and the sample gas line can be connected to the unit.
Using the quick disconnect fittings supplied, connect the vent line
followed by the sample line to the rear panel. See Figure 2-1.

Once the sample gas is flowing, set the flow rate to 0.2-2.5 SCFH.

2.6 External Signal

A 0-1 VDC output signal is included with the Model 3110. When

installed, the output signal represents the percentage of oxygen in the
current range. For instance, if the range was set for 0 to 10 ppm, then 0.1