beautypg.com

Teledyne 3110 – Portable Trace/Percent Oxygen Analyzer, FM-approved configuration User Manual

Page 36

background image

Maintenance Model

3110

Teledyne

Analytical

Instruments

28

4.4 Battery Power Supply Service

The Model 3110 is designed to be intrinsically safe, and is designed

for use only when not connected to the AC power line. TAI suggests
that an overnight recharge be performed every few days for continuous
use.

The low battery LED will begin to blink and the display will flicker

when the batteries are getting low. At this point, the batteries should be
recharged. To recharge the batteries, turn the instrument OFF (see
Section 3.2.12. Remove the instrument to a safe non-hazardous location.

Note: Depending on the application, purging may be required

before disconnecting the analyzer and moving it to a safe
location.

Plug the AC charger into a suitable 100-240 VAC 50/60 Hz power

source. Plug the other end of the charger cable into the port on the rear
of the instrument. Figure 4-1 shows the AC charger port on the rear
panel.

Note: During charging, the green battery charging indicator LED

will illuminate.

When recharging is completed, unplug the unit from the AC outlet

and disconnect the cable from the AC charger port.

Note: The analyzer cannot be turned on while the AC charger is

attached.