Teledyne 3000PAEU - General purpose percent oxygen analyzer User Manual

Page 73

background image

A-9

Percent Oxygen Analyzer

Appendix

Teledyne Analytical Instruments

to toggle between

AUTO

and

MAN

zero settling. Stop when

MAN

appears,

blinking, on the display.

Zero: Settling: Man

To Begin

Press

Enter

to begin the zero calibration. After a few seconds the first

of five zeroing screens appears. The number in the upper left hand corner is
the first-stage zero offset. The microprocessor samples the output at a prede-
termined rate. It calculates the differences between successive samplings and
displays the rate of change as

Slope=

a value in parts per million per second

(ppm/s).

#### % Zero

Slope=#### ppm/s

NOTE: It takes several seconds for the true Slope value to display. Wait about

10 seconds. Then, wait until Slope is sufficiently close to zero before
pressing

Enter to finish zeroing .

Generally, you have a good zero when

Slope

is less than 0.05 ppm/s

for about 30 seconds. When

Slope

is close enough to zero, press

Enter

. In a

few seconds, the screen will update.

Once span settling completes, the information is stored in the

microprocessor, and the instrument automatically returns to the

Analyze

mode.

NOTE: