Teledyne 3000PAEU - General purpose percent oxygen analyzer User Manual

Page 33

background image

Percent Oxygen Analyzer

Installation 3

3-11

Teledyne Analytical Instruments

st

Toggling input. Stops/Starts any status message output from
the RS-232, until st is sent again.

The RS-232 protocol allows some flexibility in its implementation.

Table 3-2 lists certain RS-232 values that are required by the 3000PA
implementation.

Table 3-2: Required RS-232 Options

Parameter

Setting

Baud

2400

Byte

8 bits

Parity

none

Stop Bits

1

Message Interval

2 seconds

3.4

Installing the Micro-Fuel Cell

The Micro-Fuel Cell is not installed in the cell block when the

instrument is shipped. It must be installed before the analyzer is placed in
service.

Once it is expended, or if the cell is exposed to air for too long, the

Micro-Fuel Cell will need to be replaced. The cell could also require replace-
ment if the instrument has been idle for too long.

When the micro-Fuel Cell needs to be installed or replaced, follow the

procedures in chapter 5, Maintenance, for removing and installing cells.

3.5

Testing the System

Before plugging the instrument into the power source:

Check the integrity and accuracy of the gas connections. Make
sure there are no leaks.

Check the integrity and accuracy of the electrical connections.
Make sure there are no exposed conductors.

Check that sample pressure is between 3 and 40 psig, according
to the requirements of your process.

Power up the system, and test it by performing the following

operations:

1. Repeat the Self-Diagnostic Test as described in chapter 4, section

4.3.5.