Teledyne 3000PAEU - General purpose percent oxygen analyzer User Manual
Page 48
4 Operation
4 Operation
4 Operation
4 Operation
4 Operation
Model 3000PA
4-14
Teledyne Analytical Instruments
If you are using password protection, you will need to enter your
password to access the alarm functions. Follow the instructions in section
4.3.3 to enter your password. Once you have clearance to proceed, enter the
Alarm
function.
Press the
Alarm
button on the front panel to enter the
Alarm
function.
Make sure that
AL1 is blinking.
AL1 AL2
Choose Alarm
Set up alarm 1 by moving the blinking over to
AL1 using the < >
arrow keys. Then press
Enter
to move to the next screen.
AL1 1.00 % HI
DftN FsN LtchN
Five parameters can be changed on this screen:
•
Value of the alarm setpoint, AL–1 #### % (oxygen)
•
Out-of-range direction, HI or LO
•
Defeated? Dft–Y/N (Yes/No)
•
Failsafe? Fs–Y/N (Yes/No)
•
Latching? Ltch–Y/N (Yes/No).
•
To define the setpoint, use the < > arrow keys to move the
blinking over to AL–1 ####. Then use the
∆∇
arrow keys to
change the number. Holding down the key speeds up the
incrementing or decrementing. (Remember, the setpoint units are
percent-of-oxygen.)
•
To set the other parameters use the < > arrow keys to move the
blinking over to the desired parameter. Then use the
∆∇
arrow
keys to change the parameter.
•
Once the parameters for alarm 1 have been set, press Alarms
again, and repeat this procedure for alarm 2 (AL–2).
•
To reset a latched alarm, go to
Dft and then press either
∆
two
times or
∇
two times. (Toggle it to
Y and then back to N.)
–OR –
Go to
Ltch and then press either
∆
two times or
∇
two times.
(Toggle it to
N and back to Y.)