Press

Escape

to re-enter the

Analyze

mode using the fixed range.

NOTE:When performing analysis on a fixed range, if the oxygen

concentration rises above the upper limit (or default value) as
established by the operator for that particular range, the
output saturates at 1 V dc. However, the digital readout and
the RS-232 output continue to read the true value of the oxy-
gen concentration regardless of the analog output range.

4.7

The

Analyze

Function

Normally, all of the functions automatically switch back to the

Analyze

function when they have completed their assigned operations. Pressing the

Escape

button in many cases also switches the analyzer back to the

Ana-

lyze

function. Alternatively, you can press the

Analyze

button at any time

to return to analyzing your sample.

4.8

Signal Output

The standard Model 3000PA Percent Oxygen Analyzer is equipped

with two 0–1 V dc analog output terminals accessible on the back panel (one
concentration and one range ID). The –MA option also has two isolated 4–
20 mA dc current outputs (one concentration and one range ID).

See Rear Panel in Chapter 3, Installation, for illustration.

The signal output for concentration is linear over the currently selected

analysis range. For example, if the analyzer is set on range that was defined
as 0–10 % O

2

, then the output would be:

Voltage Signal

Current Signal

% O

2

Output (V dc)

Output (mA dc)

0

0.0

4.0

1

0.1

5.6

2

0.2

7.2

3

0.3

8.8

4

0.4

10.4

5

0.5

12.0

6

0.6

13.6

7

0.7

15.2

8

0.8

16.8

9

0.9

18.4

10

1.0

20.0