Teledyne 2240 - Portable Handheld Hydrogen Leak Detector, 3rd generation (revision 2/29/08) User Manual

Page 27

background image

Handheld Hydrogen Leak Detector

Field Verify/Calibration

Teledyne

Analytical

Instruments

17

Step

Display

User Response

16

Apply 0.100% H2
Settle

Checking sensor temperature.

17

Apply 0.100% H2
Wait xxxx

Wait for sensor reading to stabilize, until xxxx = 0.

18

Apply 0.100% H2
Finding Average

Measuring sensor response to test gas.

19

Apply 2.000%H2
Are you sure?

With the Calibration Cup, apply 2.0% hydrogen
(balance air) to the Probe sensor. The Probe Tip
LED will remain Red. Three yellow LEDs in the LED
Bar Graph Array will turn on. Press ENTER.

20

Apply 2.00% H2
In Progress

2.00% Verify Test starts

21

Apply 2.00% H2
Settle

Checking sensor temperature.

22

Apply 2.00% H2
Wait xxxx

Wait for sensor reading to stabilize, until xxxx = 0.

23

Apply 2.00% H2
Finding Average

Measuring sensor response to test gas.

24

Enter Date:
1.0000 M

Enter the current month (1-12) using the▲ (up arrow)
and▼ (down arrow) keys.

25

Enter Date:
1.0000 D

Enter the current day (1-31) using the▲ (up arrow)
and▼ (down arrow) keys.

26

Enter Date:
6.0000 Y

Enter the current year (7=2007, 12=2012, etc.)
using the▲ (up arrow) and▼ (down arrow) keys.

27

Verify Sensor
Passed

Verify is complete, press any key.

Note the following:
1. If the unit passes verification, calibration is not required at this

time. The Verify Function stops at this point.

2. If the unit fails verification, TAI recommends calibration.

Continue with the following steps.