beautypg.com

Teledyne 2240 - Portable Handheld Hydrogen Leak Detector, 3rd generation (revision 2/29/08) User Manual

Page 21

background image

Handheld Hydrogen Leak Detector

Startup and Operation

Teledyne

Analytical

Instruments

11

Sensor Error – Unit has detected a failure to communicate with

the Probe.

broken – Unit has detected a unit internal failure
See Section 5.4 for troubleshooting

2.9 Reset Operation

The Reset Operation is used to speed recovery from hydrogen

exposure. This operation can be invoked from the keypad during Normal
Operation by Pressing and holding ◄ ► (left and right arrow buttons
simultaneously) or from the Configuration Menu (see Section 2.11).
Once invoked the user is asked to confirm the operation by pressing the
ENTER key, pressing any other key will abort the operation.

Note: The instrument must be in a hydrogen free environment

with the LCD indicating less than 0.1% hydrogen before
invoking the Reset operation.

During Reset the LCD indicates a countdown to completion and the

Probe tip LED will be yellow.

2.10 Zero Operation

The Zero Operation is used to zero the hydrogen reading if the

instrument is reporting low levels (0.001% to 0.01%) of hydrogen when
no hydrogen is present. This operation can be invoked from the keypad
during Normal Operation by pressing and holding ◄ (left arrow
button) or from the Configuration Menu (see Section 2.11). Once
invoked, the user is asked to confirm the operation by pressing the
ENTER key, pressing any other key will abort the operation.

Note: The instrument must be in a hydrogen free environment

with the LCD indicating less than 0.1% hydrogen before
invoking the Zero operation.

2.11 Configuration Menu

The Configuration Menu is used to display and modify instrument

settings and can be entered by pressing and holding the ENTER button.
The configuration parameters are summarized below with details in the